High Impact Factor : 4.396 icon | Submit Manuscript Online icon |

Pressure swing adsorption a Cleaner techniques to reduce emission

Author(s):

Prof. R. V. Prajapati , Government Engineering College, Valsad; Anant J. Songade, Government Engineering College, Valsad

Keywords:

medical oxygen concentrator, pressure swing adsorption, vacuum swing adsorption.

Abstract

Reduction of the absorber size of a medical oxygen concentrator (MOC) employing a generic pressure swing adsorption (PSA) technology is an ongoing research and development activity. The MOC typically produces a 90_93% O2-enriched product gas from ambient air at a rate of e10 L/minute (LPM) for individual use. Adsorbent columns packed with very small adsorbent particles are used to enhance the adsorption kinetics for rapid PSA cycle operation. An experimental mini-PSA set up was used to measure the performance of a simulated Skarstrom-like PSA cycle using a commercial sample of LiX Zeolite as the air separation sorbent

Other Details

Paper ID: IJSRDV1I3071
Published in: Volume : 1, Issue : 3
Publication Date: 01/06/2013
Page(s): 677-680

Article Preview

Download Article